Modular AC-powered neurosurgical power system for routine & emergency craniotomy, hematoma evacuation and transnasal skull base precise milling.
  • XGZ-ZX-Neurosurgery,Shinvaunited,XGZ-ZX-Neurosurgery

Description

XGZ-ZX neurosurgery surgical power system (Article No. HD01002) is a dedicated AC-driven surgical power tool for neurosurgery. It consists of ZX-A low speed handpiece and ZX-B high speed A handpiece, compatible with self-stopping drills, straight handpieces and milling handpieces.
The system is designed for routine craniotomy, emergency craniotomy, cerebral hemorrhage, hematoma removal, tumor resection and transnasal skull base surgery, for precise drilling and milling of intracranial and skull base bone and soft tissue.

Key Features & Clinical Benefits

High-speed & Efficient Craniotomy

• Max speed of high speed A handpiece up to 70000 rpm with 1:1 rate output

• Import fully sealed motor, power over 150W, stable high torque & high speed performance

• Supports up to 1000 sterilization cycles for long-term repeated clinical use

Intelligent Constant Power Compensation

• Constant output control with power compensation function

• Load speed drop ≤5%, maintains stable cutting performance under variable load

• Three gears with stepless speed change for flexible speed adjustment

Reliable Intraoperative Safety

• Built-in self-diagnosis on power-on; automatic stop and alarm upon fault detection

• Self-stopping drill available in Φ11mm, Φ7mm, Φ6mm options to avoid over-penetration

Comfortable Continuous Operation

• Air-free cooling and thermal insulation design

• Surface temperature rise ≤15℃ after 10 mins continuous operation, protects surgeon’s hand

Specifications

Parameter Specification
Product Model XGZ-ZX
Article No. HD01002
Application Routine craniotomy, emergency craniotomy, cerebral hemorrhage, hematoma and other neurosurgery operations
Power Source AC 220V, 50Hz
Max Rotation Speed – Low speed handpiece 1400 rpm
Max Rotation Speed – High speed A handpiece 70000 rpm
Output Torque – Low speed handpiece ≥750 mN·m
Output Torque – High speed A handpiece ≥11.2 mN·m
Speed Adjustment Three gears, stepless speed change
Output Power 300VA
Self-stopping Drill Sizes Φ11mm, Φ7mm, Φ6mm
Sterilization Supports 1000 sterilization cycles
Temperature Control Surface temperature rise ≤15℃ after 10min continuous running
Speed Stability Load speed drop ≤5%
Safety Function Power-on self-diagnosis, auto stop & alarm for faults

 